Your Position: Home > Product > Tube Furnace

1200-1400℃ Vertical Tube Furnace

 
  • Order Online
Introduction

CE Standard 1400℃ Vertical Tube Furnace

H-14SVT vertical tube furnace is designed for quenching sample from high temperature into ice water or oil rapidly to study material phase transition.

Specifications

 Ⅰ. Technical Specifications

Max. temperature

1400 ℃ (<2hour)

Working temperature

1300℃

Standard Temperature controlling

●  PID automatic control via SCR (Silicon Controlled Rectifier) power control with phase angle fired, current limiting resistor.

●  51 programmable segments for precise control of heating rate, cooling rate and dwell time.

●  Built in PID Auto-Tune function with overheating & broken Thermocouple broken protection.

●  Over temperature protection and alarm allows for operation without attendant

  

Optional of Temperature controlling

●  Software (Furnace can be operated by PC by installing a control software )

●  Touch screen temperature controller

Temperature accuracy

±1℃

Heating rate

<15 ℃ /min

Thermocouple

S type (Pt-Rh to Pt-Rh) with 99.7% purity Alumina tube

Heating element

SIC heating elements

Furnace Chamber

Three layers of ceramic fiber insulation materials

●  First layer: 1700℃ ceramic fiber board.

●  Second layer: 1430℃ ceramic fiber board.

●  Third layer: 1260℃ ceramic fiber board.

Furnace structure

Double layer steel casing with dual cooling fan, surface temperature below 60℃

Vacuum system

●  Double stage rotary vane vacuum pump with magnetic valve to prevent the oil absorbed to tube.

●  Self-expansion type double silicon ring flange, 304 stainless steel HCP.

●  Needle valve + Mechanical vacuum gauge + Discharge valve

Warranty

●  One year limited warranty with life time support. (Consumable parts such as heating elements and crucibles are not covered by the warranty, please order the replacement at related products)

●  ATTENTION: Any damages caused by the use of corrosive and acidic gases are not under the coverage of One Year Limited Warranty.

Certifications

 

Ⅱ . Application Notes

Using Attentions

●  Cooling rate shall also not exceed 10℃ /min.   

●  Toxic or explosive gases are not recommended for use with this furnace without necessary safety controls and supervision.   

●  Small cracks may appear on the surface of the refractory ceramics over extended use. This is a normal occurrence and the cracks may be repaired with alumina coating.

●  Refractory door block must be inserted before closing door.

 

Standard Size

 

Type

Max.Temperature

(℃)

Furnace Chamber Size(mm)

(Diameter*Heating zone)

Constant temperature zone(mm)

Rated Power (kw)

H-14SVT-40/300

1400

40

300

100

3

H-14SVT-60/300

1400

60

300

100

4

H-14SVT-80/300

1400

80

300

100

5

H-14SVT-100/300

1400

100

300

100

7

H-14SVT-120/300

1400

120

300

100

9

H-14SVT-40/450

1400

40

450

150

4.5

H-14SVT-60/450

1400

60

450

150

6

H-14SVT-80/450

1400

80

450

150

7.5

H-14SVT-100/450

1400

100

450

150

9

H-14SVT-120/450

1400

120

450

150

10

Note